Installation and placement conditions of cooling tower of thermal shock tester
Let's learn about the installation and placement conditions of the cooling tower of the thermal shock tester with the editor of okmarts.com!
1. The ground is flat and well ventilated;
2, there is no strong vibration around the equipment;
3, there is no strong electromagnetic field around the equipment;
4. There are no flammable, explosive, corrosive substances and dust around the thermal shock tester;
5. There is a drainage floor drain near the equipment;
6. Ground load-bearing capacity of the site: not less than 800kg/㎡;
7. Environmental conditions: temperature: 5℃~35℃; relative humidity: ≤ 85%; air pressure: 86kpa~106kpa;
8. Cooling water temperature: circulating cooling water temperature: ≤ 30℃;
9. Cooling water pressure: circulating cooling water supply pressure: 0.25MPa~0.4MPa; the design and construction of the cooling water pipeline system should ensure that the inlet pressure of the refrigerator is 0.25MPa~0.4MPa under the rated flow; the outlet of the refrigerator is cooled The pressure drop of the water tower is not more than 0.05 MPa;
10, flow rate: 7t/h, (if the circulating chilled water of the air conditioning system is used, please specify in advance);
11. Air source (compressed air): Pressure: (5~7)kg/c㎡; Connecting pipe diameter: ∮8mm; Flow rate: 0.1 cubic/min;
12. Requirements for storage environment: When the equipment is not working, the temperature of the environment should be kept within 0℃~45℃. When the environment temperature is lower than 0℃, the water stored in the thermal shock tester should be drained cleanly to avoid pipelines. The water inside freezes and damages the pipe.

Related Info
How to prevent high temperature alarm of industrial chiller in summer How to prevent high temperature alarm of industrial chiller How to prevent moisture caused by steam penetration in cold storage projects How to use small cold storage in summer with good energy saving effect Influencing factors and measures of energy saving in air conditioning and refrigeration systemsInstallation and placement conditions of cooling tower of thermal shock tester
Subscribe to relevant news content?
Subscription
Unsubscription
Recent related posts
Customers frequently viewed
IWAYA DENKI Water Pump 25CJT0752
US$2307.69
Panasonic Servo Motor MHMF022L1V2M
US$653.85
Honeywell Solenoid Valve VE4025B1045
US$334.62
ROTEN Mechanical Seal 8E5K-22-X
US$384.62
AirTAC Air Cylinder SC200X210S
US$355.00
Copeland Compressor 6DNNR37ME-TSE-250
Contact Us
Lenze Servo Drive E84AVTWC1534VX0
Contact Us
Emerson Alco Expansion Valve TER35HW35
US$538.46
CARLO Relay RZ3A60A25
US$271.40