Installation and placement conditions of cooling tower of thermal shock tester

Installation and placement conditions of cooling tower of thermal shock tester

Let's learn about the installation and placement conditions of the cooling tower of the thermal shock tester with the editor of okmarts.com!

1. The ground is flat and well ventilated;

  2, there is no strong vibration around the equipment;

  3, there is no strong electromagnetic field around the equipment;

  4. There are no flammable, explosive, corrosive substances and dust around the thermal shock tester;

   5. There is a drainage floor drain near the equipment;

  6. ​​Ground load-bearing capacity of the site: not less than 800kg/㎡;

  7. Environmental conditions: temperature: 5℃~35℃; relative humidity: ≤ 85%; air pressure: 86kpa~106kpa;

   8. Cooling water temperature: circulating cooling water temperature: ≤ 30℃;

9. Cooling water pressure: circulating cooling water supply pressure: 0.25MPa~0.4MPa; the design and construction of the cooling water pipeline system should ensure that the inlet pressure of the refrigerator is 0.25MPa~0.4MPa under the rated flow; the outlet of the refrigerator is cooled The pressure drop of the water tower is not more than 0.05 MPa;

  10, flow rate: 7t/h, (if the circulating chilled water of the air conditioning system is used, please specify in advance);

  11. Air source (compressed air): Pressure: (5~7)kg/c㎡; Connecting pipe diameter: ∮8mm; Flow rate: 0.1 cubic/min;

12. Requirements for storage environment: When the equipment is not working, the temperature of the environment should be kept within 0℃~45℃. When the environment temperature is lower than 0℃, the water stored in the thermal shock tester should be drained cleanly to avoid pipelines. The water inside freezes and damages the pipe.